Consumables & Accessories

Detector Deuterium Lamps

Replacement deuterium lamps for mainstream UV / DAD detectors from Agilent, Waters, Shimadzu, Hitachi and Thermo. High-purity quartz envelope, deuterium purity ≥ 99.999%, rated life ≥ 2000 hours, drop-in replacement with no adjustment required.

Consumables & Accessories

Detector Deuterium Lamps

Replacement deuterium lamps for mainstream UV / DAD detectors from Agilent, Waters, Shimadzu, Hitachi and Thermo. High-purity quartz envelope, deuterium purity ≥ 99.999%, rated life ≥ 2000 hours, drop-in replacement with no adjustment required.

Product Details
Replacement Deuterium Lamps for HPLC Detectors
Inquiry

Replacement deuterium lamps compatible with UV/DAD detectors from Agilent, Waters, Shimadzu, Hitachi, Thermo and other major brands. High-purity quartz envelope, deuterium fill purity ≥ 99.999%, rated life ≥ 2000 hours, stable output and low baseline noise. Direct replacement for OEM lamps, no adjustment required.

Compatible with major brandsRated life ≥ 2000 hDrop-in replacementHigh-purity quartz envelope
Rated life
≥ 2000 h
Deuterium purity
≥ 99.999%
Wavelength range
190 – 400 nm

← Swipe to see the full table →

BrandCompatible models (examples)
Agilent1100 / 1200 / 1260 / 1290 series DAD
Waters2487 / 2489 / 2998 PDA
ShimadzuSPD-M20A / SPD-M30A
HitachiL-2450 / L-7400 series
ThermoUltimate 3000 DAD
Replacement Criteria

Don't replace on burn hours alone

Rated life is a statistical figure; actual life depends heavily on the number of on/off cycles, power stability and flow cell cleanliness. Prepare to replace the lamp when any one of the following four applies:

← Swipe to see the full table →

CriterionSymptomNotes
Lamp energy dropIntensity / Energy in the self-test report falls below the instrument thresholdThe most direct indicator; most DADs show an energy curve on the self-test page
Rising baseline noiseBaseline noise clearly above the documented value under the same methodUsually appears before the energy alarm — the earliest observable signal
Loss of low-wavelength sensitivityResponse degrades markedly at 190–220 nm while higher wavelengths remain normalLamp ageing shows first at the low-wavelength end; methods detecting at 210 nm are the most sensitive to it
Cumulative burn hoursApproaching rated lifeA backstop, not the sole criterion
How to make a lamp last: lamp life is consumed mainly by burn time, but frequent on/off cycling shortens it too. Leave the lamp on through short breaks within a day and switch it off only for long idle periods (overnight or more) — this is more economical than turning it off after every run.
Replacement

Two things you must do after replacing a lamp

  1. Wavelength / lamp calibration— a deuterium lamp has two characteristic emission lines at 486.0 nm and 656.1 nm, which the instrument uses to calibrate its wavelength axis. Without calibration after replacement, wavelength accuracy cannot be assured and method validation will fail.
  2. Adequate warm-up— allow 30–60 min for baseline drift to settle before injecting. Running a sequence immediately after ignition means the drift in the first few injections gets misread as a method problem.

Handling note: never touch the quartz window with bare hands during installation — fingerprints carbonize at high temperature and permanently reduce transmission. Record the replacement date and instrument serial number and reset the burn-hour counter, otherwise the next assessment has no baseline to compare against.

Need a quote for this consumable?

Send us the specification and quantity; we quote within 24 hours