Basic Support

Analytical ColumnSelection Guide

Work through separation mode → stationary phase → substrate and pH → pore size → particle size and column dimensions, then check against mobile phase and detector, to cut down on trial-and-error with columns.

Technical support · illustrated manual

Analytical Column Selection Guide

A complete decision path from sample properties to a specific grade. Includes a separation mode decision table, a stationary phase selectivity map, pore size versus molecular weight bands, van Deemter curves and the particle size trade-off, retention versus pH, detector constraints, a quick reference for nine application areas, pharmacopoeial correspondence and the link to preparative scale-up, with calculation methods and a source index.

Version V1.0For HPLC / UHPLC analytical column selectionIssued 2026-08-0418-item source index
The suggested order is: separation mode → stationary phase chemistry → substrate and pH tolerance → pore size → particle size and column dimensions → check against mobile phase and detector.Get the separation mode wrong and later parameter adjustments generally cannot compensate; particle size and column length, by contrast, can be fine-tuned late in method development. Working in this order reduces the number of columns tried.

1. Conditions to Define Before Selection SELECTION INPUTS

Selection is a constrained decision. Before asking "which column", the six conditions below should be established; where they are unclear, any selection can only be tentative.

ConditionWhat needs to be establishedEffect on selection
Sample propertiesMolecular weight, polarity (logP), acid/base character (pKa), whether charged, whether it contains a chiral centre, solubilityDetermines the separation mode and stationary phase chemistry
Purpose of the analysisAssay / related substances / impurity identification / stability study / pre-preparative assessmentDetermines the efficiency required, the sample load and the column length
Matrix complexityPlasma, fermentation broth, food extract or purified intermediateDetermines whether a guard column is needed and how much contamination tolerance is required
Instrument conditionsSystem pressure limit, column oven range, detector type, system dead volumeDetermines the minimum particle size and minimum column bore
Method statusIn-house method / pharmacopoeial method / customer-specified method / to be filed for registrationDetermines whether the column must be chosen by USP L-number and whether a factory test report is needed
What comes nextWhether it will be scaled to semi-preparative or preparativeDetermines whether to choose a series that also offers the same phase in preparative grades
Where measured pKa and logP are not available, they can be estimated from the structure. Estimates are acceptable for initial screening, but the final mobile phase pH should be set from measured retention behaviour (see Chapter 7).

2. Selecting the Separation Mode SEPARATION MODE

The separation mode is determined jointly by the sample's solubility, polarity and molecular size. The figure below lists the corresponding modes and common stationary phases by decision criterion. Most small-molecule drug analysis falls in the reversed-phase range, but polar compounds, ionic compounds and macromolecules each have a more suitable mode.

Decision criterion Separation Modes Common stationary phases Moderately polar / hydrophobic logP > 0, soluble or partly soluble in water Reversed phase (RP-HPLC) C18 / C8 / C4 Phenyl / PFP / polar-embedded Strongly polar / no retention in reversed phase Sugars, nucleosides, polar metabolites HILIC Amide / diol / silica Zwitterionic Charged / ionizable Organic acids, inorganic anions, proteins Ion exchange (IEX) / ion chromatography Strong/weak cation and anion exchange Ligand exchange (carbohydrate analysis) Separation by molecular size Polymer distribution, protein aggregates Size exclusion (SEC / GPC) Aqueous / organic SEC Choose pore size by separation range Enantiomers / stereoisomers Chiral drugs, amino acid configuration Chiral chromatography Polysaccharide coated / bonded Crown ether, ligand exchange, derivatization Fat-soluble / insoluble in water Oils and fats, isomers, crude natural products Normal phase (NP) / preparative flash Bare silica / cyano / amino Diol
Fig. 1 Separation mode decision chart (criterion → mode → common stationary phases)
Note: the logP and solubility ranges in the figure are general guidance; borderline cases must be confirmed by measurement.Compiled for this page

The same sample can sometimes be run in more than one mode. Relatively polar nucleosides, for example, can be run on HILIC, on a polar-embedded phase, or on an aqueous-stable reversed-phase column. The choice then usually rests on detector compatibility (the high organic content of HILIC favours MS response), consistency with an existing method, and equilibration time (HILIC generally equilibrates more slowly than reversed phase).

3. Reversed-Phase Stationary Phases & Selectivity STATIONARY PHASE

Once reversed phase is settled, the choice of stationary phase mainly affects two things:retention strengthandSelectivity. Retention strength determines whether components fall in a suitable retention factor range (generally k = 2–10), and selectivity determines whether adjacent components can be separated. Chain length (C18 / C8 / C4) mainly adjusts retention strength, while aromatic, fluorinated and polar-embedded phases are used to change selectivity.

Hydrophobic retention (carbon load · chain length) Polar · aromatic · dipole selectivity C4 Macromolecules / peptides C8 Moderate retention, short run time C18 General purpose, covering most applications Polar-embedded / aqueous-stable No collapse at high aqueous content, improved peak shape for basic compounds Phenyl π–π interaction, aromatic structural analogues Pentafluorophenyl (PFP) Halogenated compounds, positional isomers, basic compounds Cyano (CN) Usable in both normal and reversed phase Diol Amino (NH₂) HILIC (amide / zwitterionic)
Fig. 2 Relative positioning of common stationary phases (schematic)
Note: the coordinates indicate relative relationships, not measured parameters; the same class of phase differs between manufacturers' products.Compiled for this page
Stationary phaseRetention characteristicsWhen to Use This ServicePoints to note
C18(ODS)Strong retention, high carbon loadMost small-molecule drugs, impurity profiling, food and environmental analysisSome products risk phase collapse under high aqueous conditions; choose an aqueous-stable grade
C8Retention about 1/2 to 2/3 that of C18Strongly hydrophobic compounds, shortening run timeSelectivity close to C18, so of limited use for improving resolution
C4Weak retentionPeptides, proteins, macromoleculesRetention is easily insufficient for small molecules
Phenyl / biphenylπ–π interaction gives aromatic selectivityAromatic structural analogues, positional isomersSelectivity weakens with acetonitrile-rich mobile phases; more pronounced in methanol systems
Pentafluorophenyl (PFP)Dipole, π–π and steric effects togetherHalogenated compounds, polar isomers, basic compoundsSelectivity is fairly sensitive to mobile phase composition; re-verify on method transfer
Polar-embedded / aqueous-stableA hydration layer forms at the amide, carbamate or other polar group100% aqueous conditions, polar compounds, improved peak shape for basic compoundsHydrophobic retention is slightly below a conventional C18 of the same carbon number
Cyano (CN)Intermediate polarity, usable in both normal and reversed phaseRapid screening, normal-phase separationBatch reproducibility is generally lower than C18; keep this in view during method validation
Amino (NH₂) / diolPredominantly polar interactionCarbohydrates, normal-phase separation, HILIC modeAmino phases can react with reducing sugars, aldehydes and ketones, giving a shorter life

Phase classes and the grades available on this site

The table maps the phase classes above onto our three product lines:COSMOSIL is the Nacalai Tesque analytical column range, with a fairly complete set of special selectivity phases;HPLCONE is our own brand, built on SilicaOne custom-bonded silica, continuous from analytical to preparative and labelled with USP L-numbers;Daisogel is Osaka Soda bulk spherical silica, for self-packed columns and industrial preparative work.

Phase classCOSMOSIL (analytical)HPLCONE (analytical / preparative)Daisogel (bulk silica)
General-purpose C185C18-MS-II (fully endcapped), 5C18-AR-II (partly endcapped, complementary selectivity)5C18A (300 m²/g surface area, high retention and high loading)SP-120-ODS-RPS (17% carbon), SP-100-ODS-HP (24% carbon)
Acid- and base-stable C185C18C (pH 1.5–12, regenerable with NaOH)SP-120-ODS-BIO (20% carbon, alkali regenerable)
Hydrophilic / aqueous-stable C185C18-PAQ5C18D (hydrophilic, for LC-MS), Ami C18 (amide polar-embedded)SP-60-ODS-RPS (60 Å, 450 m²/g surface area)
C85C8-MSC8 seriesSP-120-C8-P、SP-200-C8-BIO
C4 / macromolecule reversed phase5C4-MS、5C18-AR-300、5C8-AR-300(300 Å)SP-300-C4-BIO (3% carbon), SP-300-ODS-BIO (8% carbon)
Phenyl / aromatic selectivity5PE-MS (phenyl), πNAP (naphthylethyl), 5PYE (pyrenylethyl), PBr (pentabromophenyl), PFP (pentafluorophenyl)PE phenyl (USP L11)SP-120-C4Ph-HP (phenyl)
Shape selectivityCholester (cholesteryl), 5C22-AR-II (C22)
Cyano (CN)5CN-MS
Amino / carbohydratesNH PLUS (polymer-based amino, USP L8), SUGAR-Ca / SUGAR-HSP-120-APS-P (aminopropyl)
HILICHILICHILIC(USP L114 / L122)
Silica / normal phase5SL-IISIL (USP L3, 320 m²/g surface area)SP-120-P, SP-1000-P and other pore sizes
ChiralCHiRAL 3A / 3B / 3C、5A / 5B / 5C
Grades and parameters are taken from the product pages on this site:COSMOSIL columns · HPLCONE Columns · Daisogel silica mediaSite data
When improving resolution, changing the stationary phase usually does more than adjusting the gradient. If two components co-elute on C18, try a phenyl or PFP phase of different selectivity before extending the gradient further.

4. Substrate, Bonding & Endcapping SILICA & BONDING

At the same "C18", differences in peak shape and column life between products usually come from the substrate and the bonding process rather than the carbon chain itself. The four parameters below are normally given in the product documentation and are worth comparing item by item.

ParameterRangeEffect
Silica purityOrdinary silica / high-purity silica (metal impurities at ppm level)Residual metals interact with chelating compounds, causing tailing and reduced response
EndcappingFully / partially / not endcappedResidual silanols interact with basic compounds and cause tailing; non-endcapped products behave differently under high aqueous conditions
Bonding chemistryMono- / di- / trifunctionalPolyfunctional bonding gives better acid resistance; monofunctional bonding gives a more uniform layer
pH toleranceTypically 2–8; base-stable grades reach 12; hybrid or polymer substrates are wider stillDetermines the usable mobile phase pH range and cleaning options
Carbon loadAbout 7%–20% (varying with pore size and chain length)Correlates with retention strength, but should not be compared directly across pore sizes
Maximum temperatureTypically 60 °C; some products 80 °CA higher column temperature lowers viscosity and raises efficiency, but accelerates silica dissolution

Working near the pH limits, hydrolysis of the silica substrate and loss of the bonded phase are among the main routes to shortened column life. Below pH 2 the bonded phase hydrolyses and is lost; above pH 8 the rate of silica skeleton dissolution rises. If the method must run at pH 9–12, choose a base-stable bonded silica, a hybrid substrate or a polymer substrate; lowering the column temperature also slows degradation.

5. Matching Pore Size to Molecular Weight PORE SIZE

Pore size determines whether the solute can move freely in and out of the pores. When molecular size approaches the pore size, mass transfer is impeded, showing up as band broadening and loss of efficiency. As a rule the pore size should be at least three times the solute's radius of gyration. The figure below gives the usual correspondence.

Typical samples Small-molecule drugs · pesticide residues · additives Peptides · small proteins Proteins · enzymes Antibodies · nucleic acids · polymers 60 – 120 Å 100 Å is usual 150 – 300 Å 200 Å is usual 300 – 500 Å ≥ 1000 Å 100 Da 1 kDa 10 kDa 100 kDa 1 MDa Molecular weight (log scale) Rule of thumb: Pore size ≥ 3 × the solute's radius of gyration; too small a pore shows up as lower efficiency, peak tailing and reduced sample load. Surface area: as pore size increases, surface area falls and retention and sample load fall with it, so retention has to be balanced against mass transfer.
Fig. 3 The usual correspondence between pore size and molecular weight
Note: the ranges are general experience; for a specific product, follow the separation range stated by the manufacturer.Compiled for this page

Pore size bands and surface areas available on this site

As pore size increases, surface area falls, and retention and sample load fall with it. The Daisogel bulk silica pore size bands and their parameters are given below as a reference for selecting by pore size; COSMOSIL and HPLCONE analytical columns are mostly in the 90–130 Å range, with a 300 Å grade for proteins and large peptides.

Pore sizeSurface areaPore volumeODS carbon load*Suited to
60 Å450 m²/g0.75 mL/g19%(RPS)Small molecules and synthetic peptides where sample load matters
100 Å320 – 450 m²/g0.9 – 1.1 mL/g17%(P)/24%(HP)General-purpose reversed phase
120 Å300 m²/g1.0 – 1.1 mL/g17%(RPS)/20%(BIO)The workhorse band for general reversed phase
200 Å200 m²/g1.1 mL/g12%(RPS)/15%(BIO)Medium molecular weight peptides
300 Å100 m²/g0.9 mL/g8%(BIO)Proteins, large peptides, oligonucleotides
1000 Å25 m²/g0.9 mL/gMacromolecules, polymers
2000 Å15 m²/g0.8 mL/gVery large molecules
* Carbon load varies with grade (P / RPS / HP / BP / BIO / PK); the values here are taken from the grade database on this site. Source:Daisogel grade selectorSite data

Pore size selection in SEC follows different rules from reversed phase: SEC separates by pore size distribution, so the analyte must fall within the column's separation range (the linear part of the calibration curve) — bigger pores are not simply better. Samples with a wide molecular weight distribution can be run on columns of different pore size in series.

6 · Particle Size, Column Length and Backpressure GEOMETRY & EFFICIENCY

Efficiency rises as particle size falls, while backpressure rises with the inverse square of particle size. Choosing a particle size is really a balance betweenseparating power, run time and system pressure limit. The figure below shows plate height versus linear velocity (the van Deemter curve) for different particle sizes.

Mobile phase linear velocity u (mm/s) Plate height H (μm) 10203040 12345 10 μm 5 μm 3 μm
Fig. 4 Plate height versus linear velocity for different particle sizes (schematic curves)
Note: the dots mark the minimum of each curve — the smaller the particle, the lower and further right the minimum and the flatter the curve, so raising the flow rate costs less efficiency on a small-particle column. The curves are drawn as H = A + B/u + C·u with parameters of typical magnitude to show the trend, not measured values for a specific product.Compiled for this page
Particle sizeUsual column lengthsTypical backpressure (4.6 mm × 150 mm, 1.0 mL/min)Suited to
10 μm250 mmAbout 30–60 barRoutine detection, teaching, semi-preparative
5 μm150 / 250 mmAbout 80–150 barCovers most analytical work; commonly used in pharmacopoeial methods
3 μm100 / 150 mmAbout 200–350 barShorter run times, complex impurity profiles
< 2 μm50 / 100 mmUsually requires a system rated above 600 barUHPLC platform; system dead volume must be optimized to match
Approximate relationships for efficiency and backpressure N ∝ L / d_p efficiency rises with column length and falls with particle size ΔP ∝ L · u / d_p² backpressure rises with length and linear velocity, and with the inverse square of particle size Resolution Rs ∝ √N a fourfold gain in efficiency roughly doubles resolution

It follows that doubling the column length doubles efficiency but raises resolution only about 1.4-fold, while doubling backpressure and run time as well. Going from 5 μm to 3 μm at the same column length raises efficiency about 1.7-fold and backpressure about 2.8-fold. So where the system's pressure rating allows, reducing particle size is usually more efficient than lengthening the column.

Column IDUsual flow rateRelative sensitivity*Suited to
4.6 mm0.8 – 1.5 mL/min1.0 (reference)Routine analysis; the usual format in pharmacopoeial methods
3.0 mm0.4 – 0.7 mL/minAbout 2.4Saves solvent, compatible with conventional HPLC
2.1 mm0.2 – 0.4 mL/minAbout 4.8First choice for LC-MS; requires a low dead volume system
1.0 mm and below< 0.1 mL/minAbout 21Small samples; demanding on the system
* Relative sensitivity is estimated for equal injected amounts, taking peak volume as inversely proportional to column cross-sectional area, without allowing for system broadening or the limit on injection volume.Calculated for this page

7 · Mobile Phase and pH MOBILE PHASE & pH

For ionizable compounds, mobile phase pH usually affects retention more than the organic proportion does. The figure below shows the general form of retention versus pH for acidic and basic compounds.

Transition zone, acidic compound pKa ± 2 Transition zone, basic compound pKa ± 2 Mobile phase pH Reversed-phase retention strength 146891113 Acidic compound (pKa ≈ 4) Basic compound (pKa ≈ 9) Working window for conventional bonded silica, pH 2–8 Base-stable grades reach pH 12
Fig. 5 Retention versus pH for ionizable compounds (schematic)
Note: the curve shape follows the Henderson–Hasselbalch relationship; the actual curve is affected by organic proportion and ionic strength.Compiled for this page

For stable retention, the mobile phase pH should generally sit more than two units away from the analyte's pKa, so that it exists in a single form. Within pKa ± 1, retention time is sensitive to pH variation and method robustness suffers.

Buffer systemEffective pH rangeMS compatibleNotes
Phosphoric acid / phosphate1.9–3.1;6.2–8.2NoLow UV cut-off, commonly used in pharmacopoeial methods; not usable with MS or ELSD
Formic acid / ammonium formate2.8–4.8YesCommon in LC-MS, with good response in positive ion mode
Acetic acid / ammonium acetate3.8–5.8YesFairly wide applicability, usable in both positive and negative ion mode
Ammonium bicarbonate9.2–10.2YesFor MS methods under basic conditions; requires a base-stable column
Trifluoroacetic acid (TFA)About 2.0LimitedImproves peptide peak shape but suppresses ESI response; quantitative methods must assess this
Buffer concentration is usually 10–50 mmol/L. Too low and buffering capacity is insufficient, so pH shifts with each injection; too high and the risk of salting out and blockage increases, particularly with a high proportion of acetonitrile.

As for organic modifiers, acetonitrile is a stronger eluent than methanol and less viscous, so it generates less backpressure; the hydrogen bonding of methanol offers different selectivity, most noticeably on phenyl columns. During method development the two can be screened as independent variables.

8. Detector-Imposed Constraints DETECTOR CONSTRAINTS

The detector determines which mobile phase components are unusable, and this constraint generally comes before the choice of column. Discovering late in method development that the mobile phase and detector are incompatible often means selecting a new column.

DetectorMobile phase restrictionsColumn preferenceNotes
UV / DADMust be transparent at the detection wavelength; acetonitrile has a lower cut-off than methanolNo particular restrictionAt low wavelengths (< 210 nm), avoid carbonyl solvents and TFA
Fluorescence (FLD)Avoid fluorescence quenchersMatched to the derivatization schemeAmino acids detected after OPA / FMOC derivatization, with higher sensitivity than UV
Refractive index (RI)Isocratic only; column and ambient temperature must be stableDedicated carbohydrate columns, SEC columnsA gradient causes baseline drift and cannot be used
ELSD / CADThe mobile phase must be entirely volatile; phosphate is prohibitedReversed phase or HILICResponse is non-linear with concentration; quantify with a logarithmic fit or a narrow calibration range
Mass spectrometry (MS)Non-volatile salts prohibited; ion-pairing reagents suppress response2.1 mm bore columns, MS-compatible phasesAmmonium formate and ammonium acetate are the usual choices; the degree of suppression by TFA must be assessed
Conductivity / suppressedRequires a matching eluent systemDedicated ion chromatography columnsAnalysis of inorganic anions and cations

9 · Applications and Product Correspondence APPLICATION MATRIX

The table gives a usual starting point by sample class. The grades listed are current products on this site; the actual selection must still be checked against the conditions in Chapter 1.

Sample classRecommended starting pointMobile phase systemAlternatives and notes
Peptides · short peptidesHPLCONE 5C18A (high loading) / 5C18C (acid- and base-stable); Daisogel SP-120-ODS-RPS for preparative work0.1% TFA or formic acid / acetonitrile gradientFor large peptides, switch to C4 or a 300 Å pore size; at preparative scale, watch the sample load
Related substances in drugsCOSMOSIL 5C18-MS-II or HPLCONE 5C18A, 5 μm × 150 mmBuffer / acetonitrile gradientFor co-eluting components switch to phenyl or PFP; methods for filing must fix the brand and lot information
Carbohydrates · honeyHPLCONE SUGAR-Ca (calcium form) / SUGAR-H (hydrogen form)Pure water or dilute sulfuric acid, isocraticRI detection, column temperature 60–80 °C; the hydrogen form can also separate organic acids
Amino acidsHPLCONE 5C18D with pre-column derivatization (OPA / FMOC)Buffer / acetonitrile gradientConfiguration can be analysed with a chiral derivatizing reagent, without a chiral column
Nucleotides · oligosaccharidesHPLCONE 5C18D, COSMOSIL 5C18-PAQ or a HILIC phaseHigh aqueous or high acetonitrileA hydrophilic C18 retains stably in a fully aqueous mobile phase, which suits MS coupling
Polar metabolites (LC-MS)COSMOSIL HILIC or HPLCONE HILIC (zwitterionic)Acetonitrile / ammonium acetate gradientEquilibration takes longer; equilibrate fully before starting a sequence
Aromatics · polycyclic aromatic hydrocarbonsCOSMOSIL 5PYE / πNAP / PBr; HPLCONE PE phenyl for routine aromaticsMethanol / water or acetonitrile / waterπ–π interaction gives selectivity complementary to C18
Protein molecular weight distributionShodex aqueous SEC series (pore size chosen by separation range)Phosphate buffer, isocraticKeep conditions mild to preserve the native conformation; avoid non-specific adsorption on the substrate
Chiral drugsCOSMOSIL CHiRAL 3A–3C / 5A–5C; ChiralONE series for preparative workNormal phase, reversed phase or polar organic modeScreen the mobile phase mode first, then fix the phase; bonded types tolerate a wider solvent range
Grade details:COSMOSIL · HPLCONE · Daisogel · Shodex · ChiralONE · PharmacopoeiaSite data

COSMOSIL — special selectivity phases

COSMOSIL is made by Nacalai Tesque, with Microwants as distributor for mainland China. Beyond the usual C18 / C8 / C4, the range offers a number of aromatic and shape-selective phases, suited to structural analogues that C18 cannot resolve.

GradeStationary phaseMechanismTypical use
5C18-MS-IIODS, fully endcappedHydrophobicRoutine analysis, good peak shape for basic compounds
5C18-AR-IIODS, partly endcappedHydrophobic + residual silanol interactionComplementary selectivity to MS-II
5C18-PAQAqueous-stable ODSHydrophobic; no collapse at high aqueous contentPolar compounds, 100% aqueous conditions
5C22-AR-IIC22Hydrophobic + shape selectivityLong-chain and configurational isomers
CholesterCholesterylShape selectivityFat-soluble vitamins, steroids
πNAPNaphthylethylπ–πAromatic positional isomers
5PYEPyrenylethylπ–π + dispersion forcesPolycyclic aromatics, fullerene-related separations
PBrPentabromophenylHydrophobic + dispersion forcesHalogenated compounds, steroids
PFPPentafluorophenylπ–π + dipoleHalogenated compounds, polar isomers
5PE-MSPhenylπ–πAromatic compounds
5C18-AR-300 / 5C8-AR-300300 Å wide-pore ODS / C8HydrophobicProteins and large peptides
CHiRAL 3A–3C / 5A–5CPolysaccharide chiral phaseChiral recognitionEnantiomer separation
For the full range of grades, particle sizes and part numbers, see the COSMOSIL column seriesSite data

HPLCONE — our own brand, labelled with USP L-numbers

HPLCONE uses our own SilicaOne custom-bonded silica (total metal impurities < 10 ppm), with continuous analytical and preparative specifications and USP packing codes labelled on several grades, making column selection for pharmacopoeial methods straightforward.

GradeUSP codeCharacteristicsSuited to
5C18AL1300 m²/g surface area, 12 nm pores, pH 2–10, usable in 100% aqueousPeptides, antibiotics, natural products, balancing retention against sample load
5C18CL1High bonding density and full endcapping, pH 1.5–12, regenerable with NaOHSmall peptide separation, methods with demanding acid or base conditions
5C18DL1Hydrophilic bonding, no collapse in fully aqueous systems, no ion-pairing additive neededOligosaccharides, amino acids, nucleotides, organic acids, LC-MS coupling
Ami C18L60 / L1Amide polar-embedded, multi-mode interactionAcids, amines, phenols and other compounds insufficiently retained on C18
PE phenylL11π–π interactionAromatics and structural analogues
SILL3Spherical fully porous ultra-pure silica, 320 m²/g surface areaNormal-phase separation
NH PLUSL8Polymer-based bonded amino, tolerant of alkaline cleaningCarbohydrate quantification
HILICL114 / L122Zwitterionic hydrophilic phasePolar compounds, metabolomics, glycopeptides
SUGAR-Ca / SUGAR-HL19 / L17Ligand exchange, pure water mobile phase, column temperature 60–80 °CCarbohydrates in honey and food; the hydrogen form also detects organic acids
For specifications and L-number correspondence, see the HPLCONE column series and PharmacopoeiaSite data

Daisogel — bulk silica for self-packing and scale-up

Daisogel is Osaka Soda spherical silica, with Microwants as China master distributor. Grades are named as "pore size – particle size – functional group – grade", with pore sizes of 60–2000 Å and particle sizes of 3–50 μm; the same phase runs continuously from analytical to industrial preparative particle sizes, which keeps selectivity constant on scale-up.

GradeCharacteristicsSuited to
PConventional bonding, general purposeMost routine separations
RPSHigh bonding density, fairly complete endcappingThe workhorse for general reversed phase; SP-120-ODS-RPS at 17% carbon
HPHigh carbon loadWhere retention and sample load matter; SP-100-ODS-HP at 24% carbon
BPLower carbon load, slightly more polarGives selectivity different from RPS
BIOModified silica with a wider pH tolerance, regenerable with NaOHPeptide and protein purification, processes requiring alkaline regeneration
PKFlagship grade, with tighter batch consistencyRegistered processes and long-term supply
Grade combinations and parameters can be looked up in the Daisogel grade selector ; for packing and scale-up see Analytical Column Packing and DAC Packing SupportSite data

10. Pharmacopoeial Methods & Compliant Selection PHARMACOPOEIA

When running a pharmacopoeial method, the choice of column is constrained by the monograph. Pharmacopoeias specify a packing code rather than a brand, so products may be substituted within the same code, but system suitability must be demonstrated after any change.

SystemHow it is specifiedPermitted adjustment
USPSpecifies the packing class by L-number (L1 octadecylsilane bonded silica, L7 octyl, L11 phenyl, L3 silica, L10 cyano, L8 amino)Under USP <621>, in isocratic methods the column length and particle size may be adjusted so long as L/dp stays within −25% to +50%; the limits for gradient methods are stricter
ChP (Chinese Pharmacopoeia)General chapter 0512 sets the chromatographic conditions and system suitability requirementsChanges to the packing class and mobile phase composition must comply with the general chapter, and system suitability must be re-established
EPSpecified by packing descriptionPermitted adjustments are set out in general chapter 2.2.46

System suitability usually covers four items: theoretical plate number, tailing factor, resolution and repeatability. All four should be re-measured and recorded after a column change. For methods intended for regulatory filing, it is advisable to retain the brand, grade, lot number and factory test report of the column used.

The correspondence between L-numbers and current products can be looked up in the Pharmacopoeia , which lists every USP L-series code with the brands available.

11. Transition to Preparative Scale SCALE-UP

If the analytical method will later be scaled to semi-preparative or preparative work, the transition should be considered at the selection stage. The three points below, settled during the analytical phase, reduce rework on scale-up.

ItemWhat to do at the analytical stageBenefit on scale-up
Availability of the phaseChoose a phase series that also exists in 10 μm and 20 μm preparative gradesSelectivity is preserved on scale-up, with no need to redevelop the method
Scalability of the mobile phaseAvoid non-volatile salts such as phosphate that are difficult to recoverLowers the cost of downstream desalting and solvent recovery
Loading conditionsRecord the overload threshold (where injection amount and peak height depart from linearity)Scale the loading in proportion to media mass, giving a sound basis for capacity estimates
Scale-up conversion (constant linear velocity) F₂ = F₁ × (d₂ / d₁)² flow rate scales with the square of the column bore m₂ = m₁ × (d₂ / d₁)² sample load scales with cross-sectional area (bed height unchanged) t_G2 = t_G1 gradient time is unchanged; gradient volume scales automatically with flow rate

On availability of the phase, a given Daisogel pore size and functional group combination can be had at 3, 5, 10, 15, 20, 30, 40 and 50 μm, so selectivity can be confirmed at analytical scale with a small particle and the same bonded phase carried to preparative scale at a larger particle, which avoids re-screening phases on scale-up. HPLCONE analytical and preparative columns use the same SilicaOne substrate, with specifications extending from 4.6 mm bore to 20 mm and custom sizes.

Packing and acceptance requirements after scale-up are covered in Analytical Column PackingDAC Packing Support and DAC Packing Problem Diagnosis

12. Selection Self-Check List CHECKLIST

No.CheckAction if it fails
1The separation mode is consistent with the sample's polarity, molecular weight and charge stateReturn to Chapter 2 and re-decide
2k for the target components falls in the range 2–10Adjust the organic proportion; if it is still outside, change chain length
3Resolution for the critical pair is ≥ 1.5Change the stationary phase selectivity first, and adjust the gradient second
4Mobile phase pH is more than two units away from the analyte's pKaAdjust the pH, or move to ion pairing or HILIC
5Mobile phase pH is within the chosen column's toleranceSwitch to a base-stable, hybrid or polymer substrate
6Pore size matches the sample's molecular weightAdjust the pore size per Chapter 5
7Expected backpressure does not exceed 80% of the system limitIncrease particle size, shorten the column, or reduce the flow rate
8Mobile phase is compatible with the detectorChange the buffer system, and reselect the phase if necessary
9A guard column is fitted for complex matricesAdd a guard cartridge of the same phase
10For a pharmacopoeial method, the L-number and system suitability requirements have been checkedConsult Pharmacopoeia and re-measure the four criteria
11If scale-up is intended, preparative grades of the same phase are confirmed availableChange to a phase series that has preparative grades

Appendix A Calculation Methods and Assumptions METHODS & ASSUMPTIONS

The equations used on this page are set out below for checking and recalculation. Symbols: tR retention time, t0 dead time, W1/2 width at half height, L column length, dp particle size, dc column bore, ε total column porosity (taken as 0.65 for estimation), F flow rate.

Retention factor k = (t_R − t_0) / t_0 target range for selection 2 – 10 Theoretical plates N = 5.545 × (t_R / W_1/2)² usual system suitability limit N ≥ 2000 Reduced plate height h = L / (N × d_p) packing quality criterion, h ≤ 3 Resolution Rs = 2 × (t_R2 − t_R1) / (W_1 + W_2) baseline separation Rs ≥ 1.5 Column volume V_m = π × (d_c/2)² × L × ε equilibration and flush volumes are multiples of this Linear velocity u = F / (π × (d_c/2)² × ε) held constant on scale-up Gradient conversion t_G2 = t_G1 × (V_m2 / V_m1) × (F_1 / F_2) keeps the gradient volume equivalent

The backpressure ranges given are estimated for an acetonitrile/water mobile phase (about 50:50, 25 °C, viscosity about 0.8 mPa·s) and vary with mobile phase composition, column temperature and medium batch; they are for initial selection, not for acceptance. Relative sensitivity is calculated for equal injected amounts with peak volume inversely proportional to column cross-sectional area, without allowing for system broadening or the injection volume limit, so measured values are generally lower than calculated.

The pore size to molecular weight ranges are general empirical values (inferred for this page; no statistical source); the coordinates in the stationary phase map indicate relative relationships and do not represent measured parameters for any product.

Appendix B Source Index REFERENCES

This page is compiled from the public sources and existing site data below. Pharmacopoeial text follows the current edition; please check the latest version when citing.

Pharmacopoeias and regulations
[1]USP general chapter <621> Chromatography, including the L-number packing classification and permitted method adjustmentsusp.org
[2]USP general chapter <1225> Validation of Compendial Proceduresusp.org
[3]Pharmacopoeia of the People's Republic of China, 2020 edition, Part IV, general chapter 0512, High Performance Liquid Chromatographychp.org.cn
[4]European Pharmacopoeia 2.2.29 Liquid chromatography; 2.2.46 Chromatographic separation techniquesedqm.eu
[5]ICH Q2(R2) Validation of Analytical Proceduresich.org
[6]ICH Q3A(R2) / Q3B(R2) Impurities in new drug substances and new drug productsich.org
[7]ICH Q14 Analytical Procedure Developmentich.org
Monographs and methodology
[8]Snyder L.R., Kirkland J.J., Dolan J.W. Introduction to Modern Liquid Chromatography, 3rd ed., Wiley
[9]Snyder L.R., Kirkland J.J., Glajch J.L. Practical HPLC Method Development, 2nd ed., Wiley
[10]Neue U.D. HPLC Columns: Theory, Technology, and Practice, Wiley-VCH
[11]Guiochon G. et al. Fundamentals of Preparative and Nonlinear Chromatography, 2nd ed., Academic Press
[12]van Deemter J.J., Zuiderweg F.J., Klinkenberg A. Chem. Eng. Sci., 1956, 5, 271 — plate height and linear velocity
[13]Knox J.H. J. Chromatogr. Sci. — reduced parameters and packing quality criteria
[14]IUPAC Recommendations, Nomenclature for Chromatography (Pure Appl. Chem., 1993, 65, 819)iupac.org
Product and site material
[15]COSMOSIL column technical documentation (Nacalai Tesque)nacalai.co.jp
[16]Daisogel chromatographic silica product documentation (Osaka Soda DAISO)daiso-chem.co.jp
[17]Pages on this site:Pharmacopoeia section (USP L-number reference)FAQTroubleshooting ManualCalculatorsmicrowants.com
[18]Pages on this site:Analytical Column PackingDAC Packing SupportDAC Packing Problem Diagnosismicrowants.com
The parameter ranges on this page are for initial selection. For methods involved in regulatory filing, pharmacopoeial compliance or process scale-up, measured data and the current regulatory text take precedence. For selection advice on a specific sample, send us the sample properties and the purpose of the analysis and。